Microwave Digestion & Extraction System Parameter
• - Microwave output power and frequency: 0-1300W, stepless adjustment; 2450 MHz
• - Microwave résonance radiation cavity: 65L, internally installée] corrosion résistant internally clapboard
• - Multi-vessel pressure monitoring: 0.1~5MPa (precision:±0.01MPa); multiple non-contact pressure sensors, real-time monitoring, display and control the true pressure of 12 digestion vessels
• - Multi-vessel température monitoring: 0-250°C(precision:±1°C); multiple non-contact température sensors, real-time monitoring, display and control the true température of 12 digestion vessels
• - Number of digestion sample: simultaneously digest 12 samples of same or different interstitial substance contents. The user mayfreely digest or extract 1-12 samples
• - Multiple safety measures:
High pressure exteriorvessel: PSF exteriorvessel with PPS liner, maximum pressure endurance is 6MPa (900 psi). Overheating protection: when the case or the exterior vessel is overheated, instrument will warn and suspend the microwave radiation
Abnormal noise détection: when abnormal noise is made in the furnace chamber, instrument will warn and eut off microwave automatically.
Abnormal pressure warning: a unique device of optical fiber non-contact displacement sensor', real-time monitor and control the pressure in the digestion vessel, display every vessels' pressure change by screen. Independently monitor each vessel's reaction situation, avoid overpressure caused by single sample's too violent reaction.