The LEDCure™ Radiometer provides a patented “Total Measured Optic Response.” Everything in the UV LED radiometer’s “optics stack” is included in the overall instrument response, not just the optical filter. The Total Measured Optic Response provides consistent, repeatable and accurate readings run to run, unit to unit and source to source.
Easy to use format, provides irradiance (W/cm²) and energy density (J/cm²) values on the display as well as the irradiance profile
Works with EIT’s UV PowerView III or PowerView II software
Standard version shows all data on the display
Display can be set to Graph or Reference Mode
Profiler version shows data on the display plus downloads the irradiance profile (W/cm² as a function of time) to a computer for further analysis
Profiler allows easy comparison of different LEDs, power levels, height above substrate and exposure time/line speed changes
High dynamic range (40 W/cm²), user changeable battery
Available in 4 different LED-specific response bands (L-bands)
Operating Range - 400 mW/cm2–40 W/cm2
Spatial Response - Approximately cosine
Accuracy - ±10% of reading plus ±0.2% of full scale
Typical ±5% or better
Material - Aluminum & Stainless Steel
Dimensions (D x H) - 117 x 12.7 mm
Weight - 10.1 oz (289 grams)
Carrying Case - Supplied with carrying case, cut polyurethane foam interior, scuff resistant nylon exterior cover